Ultrasonic cleaning machine common fault one: transducer common fault damp
Check whether the transducer is damp, just use the MEGOHM meter to check the transducer insulation resistance can, if the resistance value does not meet the requirements, then you can be sure that the transducer is damp. The Solution: place the entire transducer (other than the spray-molded case) in an oven and bake at a high temperature of about 100 Baidu for more than three hours until the insulation resistance meets the requirements when tested with a MEGOHM meter.
Two: Energy Device Common Fault of the vibrator ignition and degumming
It is found that the vibration of the vibrator will be affected if the vibrator is ignited and the ceramic material is broken, and the vibrator will be burnt if it is not treated in time. In this case should find a technician to check, will be broken off the vibrator, so as not to affect the normal use of other vibrators. The damaged vibrator is then returned to the manufacturer for repair.
Three: stainless steel vibration surface perforation
If the transducer of ultrasonic cleaning agent is used too long at full load, the vibration surface may be perforated, which is caused by the long-time high-frequency vibration fatigue of the stainless steel plate on the vibration surface, a good solution is to purchase the same type of vibration surface replacement.